Product introduction:
1. All Rockwell and surface Rockwell scales can be measured
2. Conduct the force test according to the standard.
3. The z-axis of the worktable rises and falls automatically, and the x-axis and Y-axis directions of the
sample table move automatically.
4. Press the start key, the worktable will rise automatically, the sample will be loaded and unloaded
automatically after touching the indenter, and the hardness value will be displayed automatically;
Automatically move to the next measurement point for automatic measurement.
5. According to the standard hardness block, each hardness scale is automatically corrected in three
sections: high, medium and low.
6. Measure any point on any sample within the range of 100mm * 100mm.
7. High precision X / y automatic mobile station is adopted, and the repetition accuracy is 0.005 mm.
8. Set measuring points, conduct multi-point full-automatic testing, batch testing, and greatly improve
the efficiency.
9. Various fixtures can be customized for batch testing of different samples.
10. The database processing function is powerful, including: HR, Hb, HV and other hardness system
conversion, setting the qualified range, exceeding the limit automatic alarm, and inputting the tester,sample name and other information. The measurement parameters and data are automatically saved
to the database.
11. When the hardness tester host is used independently, the data USB flash disk is directly saved in
Excel format for easy editing and processing.
12. The product is designed and produced by itself, has independent intellectual property rights and
software copyright, and can be customized and upgraded for a long time.
Technical Parameters:
Model |
HTR-150T-XYZ-A |
HTR-150/45T-XYZ-A |
Initial test force |
10kgf(98.07N) |
10kgf(98.07N) 3.0kgf(29.4N) |
Total test force |
60kgf(588N) |
60kgf(588N) |
Hardness scale |
HRA、HRB、HRC、 |
HRA、HRB、HRC、HRD、HRE、HRF、 |
Test scope |
HRA:20-88、HRB:20- |
HRA:20-88、HRB:20-100 、HRC:20- |
standard |
GB/T230 , ISO 6508, ASTM E18, BSEN 6508 |
|
Duration Time |
0-99s |
|
Hardness value |
0.1HR |
|
Test force error |
<1.0% |
|
data output |
LCD display |
|
Exchange Scales |
Surface Rockwell, Brinell |
Rockwell, Superficial Rockwell, Brinell, |
Maximum |
300mm for ordinary sample table; 200mm with X mobile station; 100mm |
|
Distance from |
200mm |
|
X / y motor table |
100mm(optional: 200mm) |
|
Accuracy of X / y |
0.005mm |
|
size |
560×280×850mm |
|
weight |
80kg |
|
Power Supply |
AC220+5%,50~60Hz,600W |